Certified Biomedical Equipment Technician (CBET) Practice Exam 2025 – Comprehensive All-In-One Guide for Exam Success!

Question: 1 / 400

What is the test load resistance for a defibrillator analyzer?

50 ohms

The test load resistance for a defibrillator analyzer is typically set at 50 ohms. This resistance is used to simulate the electrical load presented by human tissue during defibrillation. When testing defibrillators, it is crucial that the analyzer is configured to the proper load so that the results accurately reflect the performance of the device when it is used in a clinical setting. A 50-ohm load is standard because it closely mimics the resistance encountered in actual defibrillation scenarios, ensuring that the analyzer can effectively evaluate the energy output and waveform characteristics of the defibrillator being tested. This allows biomedical equipment technicians to verify that the defibrillators are functioning correctly and delivering the intended therapeutic effects during use.

Get further explanation with Examzify DeepDiveBeta

75 ohms

100 ohms

150 ohms

Next Question

Report this question

Subscribe

Get the latest from Examzify

You can unsubscribe at any time. Read our privacy policy